Slaughter Test Systems
  English 中简 中繁  
(800) 421-1921 (USA) / (580) 223-4773
 Home Products Contact STS About STS More Info
Model 1740
Test System / Specifications


Downloads
English Brochure - PDF (2,087Kb)
Chinese Brochure - PDF (3,079Kb)

Complete System
Description Specifications
Power 115 / 230 V, 15A, 47 - 63 Hz (UPS Dependent)
Input Voltage 115 / 230 V selectable, ± 10 % variation; 47 - 63 Hz
Operating Environment Temperature: 0 - 40 Deg. C
Relative Humidity: 0 - 85%, Non-Condensing

Hipot
Description Operation
Output Rating 150 - 5,000 VAC, 0.01 - 25.00 mA, 50 or 60 Hz Programmable
150 - 6,000 VDC, 0.01 - 10.00 mA
Voltage Setting 150 - 5,000 VAC, 5 Volts per step
150 - 6,000 VDC, 5 Volts per step
Accuracy: +/- 2% of Setting + 5 V
(Relative to Displayed Output)
Dwell Time 0.1 - 999.9 Seconds, 0.1 Second Increments
Ramping 100 - 50,000 Volts per Second
Failure Settings Current - Capacitive, Resistive and Total
AC Limits - 1.0 uA - 25.00 mA, 1.0 uA Increments
DC Limits - 1.0 uA - 10.00 mA, 1.0 uA Increments
Arc Limits - 0 - 40 amps/millisecond
Display Capacitive, Resistive and Total Leakage

Surge
Description Operation
Output Rating 250 - 5,000 V Peak
Voltage Setting 250 - 5,000 V, 5 Volts per step
Accuracy: +/- 2% of Setting + 5 V
(Relative to Displayed Output)
Failure Settings High Limit: +99% Deviation
Low Limit: -99% Deviation

Armature Resistance & Weld
Description Operation
Output Rating DC: 0 - 10 V, 1.5 A Max, Regulation: +/- 1%
Failure Settings Resistance
   High Limit: 10 uΩ - 100 Ω
   Low Limit: 10 uΩ - 100 Ω
Weld
   High Limit: 10 uΩ - 99,999 uΩ
Display Resistance
   Range: 10 uΩ - 9999 Ω
   Resolution: 1 uΩ
   Accuracy: +/- 1% of Reading
Weld
   Range: 1 uΩ - 99,999 uΩ
   Accuracy: +/- 1% of Reading
IR Temperature Comp. Range: 0 - 115°C; Accuracy: +/- 4°C
Range: 115 - 350°C; Accuracy: +/- 6%
Ambient Temperature Comp. Range: 10.0 - 40.0°C
Accuracy: +/- 2°C

DC Resistance
Description Operation
Output Rating DC: 0 - 10 V, 1.5 A Max, Regulation: +/- 1%
Failure Settings Resistance
   High Limit: 100 uΩ - 400 kΩ
   Low Limit: 100 uΩ - 400 kΩ
Display Resistance
   Range: 100 uΩ - 10 MegΩ
   Resolution: 1 uΩ
   Accuracy: +/- 1% of Reading
IR Temperature Comp. Range: 0 - 115°C; Accuracy: +/- 4°C
Range: 115 - 350°C; Accuracy: +/- 6%
Ambient Temperature Comp. Range: 10.0 - 40.0°C
Accuracy: +/- 2°C

All specifications subject to change without notice.
Copyright 2004 - Slaughter Test Systems